Kanomax 3885 Condensation Nucleus Particle Counter
Condition: Brand New
The Kanomax 3885 Condensation Nucleus Particle Counter is ideal for continuous multi-point monitoring in cleanrooms, IC Production, HDD Testing, and MEMS. With 0.01µm sensitivity, less downtime by less alcohol replacement, and easy checking of the current status on the LEDs, the Kanomax 3885 provides ultrafine particle detection, easy maintenance, and no more headaches in the cleanroom, using Propylene Glycol.
Features:
- Condensation Nucleus Counter for continuous multi-point monitoring in cleanroom.
- 0.01µm sensitivity: Ideal for ultrafine particle detection.
- Less downtime by less alcohol replacement: No more headaches in cleanroom, using Propylene Glycol.
- Easy checking of the current status on the LEDs: Easy maintenance.
Operating Principle: Ultrafine airborne particle smaller than 0.1µm in clean rooms, which generally cannot be detected by ordinary light-scattering particle counters, can be optically measured by condensation growth of particle through supersaturated conditions formed by compulsion mixing of sampled air and high temperature alcohol vapor. The best feature of this model is its ability of continuous monitoring of particles. Owing to larger sampling rate than previous model, it becomes possible to measure and verify the particle level within the short time, even in lean concentration environment in clean rooms. Vacuum source is located outside, so there will be no emission of dust from the instrument.
Specifications:
- Product Name: Condensation Nucleus Counter
- Model Number: Model 3885
- Type: Sensor-Counter Isolated Type
- Principle: Vapor/Air Mixing Type Condensation Particle Counting
- Maximum Concentration: 3×106 Particles/cf (100 Particles/cc)
- Sensitivity: 0.01µm (50% Counting Efficiency)
- Measuring Accuracy: 0.02µm: 100% ± 10%
- Coincidence Loss: Less than 5% at 3,000,000 Particles/cf
- Sampling Flow Rate: 0.1CFM (2.83L/min) ± 10%
- Flow Rate: 4.2 ± 0.4L/min (Requires external vacuum source)
- Alcohol: Propylene Glycol
- Indicators: Ready, Liquid Level, Flow, Optics, Particle
- Interface: RS-232C and RS-485
- Output: Output for M/P or Analog Output
- Operating Environment: Temperature: 20~30ºC (68~86ºF) Clean Evironment (Such as Clean Room and so on)
- Dimensions: 250(W) × 250(H) × 280(D)mm (Excludes alcohol bottle)
- Weight: Approx. 10Kg (22lbs)
- Power Supply: AC100 ~ 240 V
- Standard Accessories: Includes AC Cable, Operation Manual, Alcohol Bottle (200cc), Bottle Holder
- Options: Alcohol Bottle (1000cc), Bottle Holder (for 1000cc)
|
|